Yiyuntian BTYQ-MS104KBTYQ-MS104K-S Portable Gas Detector

The Yiyuntian BTYQ-MS104KBTYQ-MS104K-S is a portable gas detector.

  • Detectable gases: Measuring range
  • Measuring range: Resolution
  • Sensor type: Electrochemical and low-power infrared, chosen to suit range, site and user requirement
  • Explosion protection: Intrinsically safe Ex ia IIC T4 Ga
SKU: GD-0305 Category: Tag: Brand:

Description

The Yiyuntian BTYQ-MS104KBTYQ-MS104K-S is a portable gas detector.

Selectable units: umol/mol, ppm, mg/m3, %Vol and %LEL.

GB 15322.3-2019 Combustible Gas Detectors – Part 3: Portable combustible gas detectors for industrial and commercial use.

Single gases such as carbon monoxide, hydrogen sulphide, oxygen, combustible gas and carbon dioxide.

Selectable units: umol/mol, ppm, mg/m3, %Vol and %LEL.

Temperature -20 to +50 C, customisable to -40 to +70 C; relative humidity 10-95 % RH.

Specifications

Brand Yiyuntian
Origin category Domestic (China)
Display units Selectable units: µmol/mol, ppm, mg/m³, %Vol, %LEL
Detection principle Electrochemical and low-power infrared, chosen to suit range, site and user requirement
Sensor life Electrochemical 2-3 years; infrared 5-10 years
Display 1.7 in LCD
Display content Live concentration, alarms, battery, charging status, units and gas name
Accuracy Typical accuracy: ≤±3 %FS (higher-accuracy sensor available)
Response time T90 ≤30 s (typical)
Power supply DC3.6V
Battery capacity 1600 mAh energy cell, no charging, 1-2 years operation
Communication interface Infrared; optional communication dock for data transfer, manual calibration and export
Explosion protection type Intrinsically safe Ex ia IIC T4 Ga
Ingress protection IP68
Housing material High-strength flame-retardant polycarbonate, fully rubberised, drop, wear and static resistant, drop rating ≥8 m
Dimensions Maximum size 87 x 63 x 32 mm (L x W x H)
Detectable gases Measuring range
Resolution Detectable gases
Measuring range Resolution
Communication interface Infrared; battery capacity 1600 mAh
Omnidirectional alarm indication / alarm method Audible, visual and vibrating alarm
Selectable display units µmol/mol, ppm, mg/m³, %Vol, %LEL; principle: electrochemical and low-power infrared
Gas name / accuracy / typical accuracy ≤±3 % FS (higher-accuracy sensor available); uncertainty ≤±2 %; linearity ≤±3 %; repeatability ≤±2 %